Chlorine detector

The main feature of the SHINE portable chlorine detector is that it can quickly detect the sample,and its test results can be used as the preliminary basis reference data for legal decisions

Technical index

Lower limit of determination:5.0mg/L

Determination range:Low range 0.0~500.0mg/L

                                    High range 0.05%~65%

Measurement accuracy:5%

Light source:LED silicon photodiode(Wavelength 510nm)


  Contact Now
Product Details

When using SHINE chlorine detection instrument, the operator should follow the instruction manual or pass the necessary operation knowledge training of our company befor using the instrument, The reagent used in the instrument is chemical,and the optimal storage temperature is 2℃~8℃,The operator must wear protective gloves and glasses when using the detection reagent, if you accidentally spill the reagent on your body,wash it with water immediately

Testing instrument


Main engine and accessories

  • Testing instrument:GDYS-104SM (1)

  • Electronic platform scale:(1)

  • Pipette:100~100μL (1)

  • Color bottle:10mL(2)

  • Extraction bottle:100mL(1)

  • Aluminum alloy Box(1)

  • Effective chlorine reagent 50tiems/set(1) Order number:4S104SM-2-1

  • File inclusion

  • Instruction Manual(1)

  • Product Acceptance form(1)

  • Product Quality certificate(1)

Instructions for using chlorine detection equipment

  • The test water must be distilled water or deionized water

  • All glassware should be cleaned with acid wash solution,tap water and distilled water in the chemical laboratory befor the first use

  • After each measurement, the solution in the colorimetric bottle should be poured out in time,and then cleaned with acid washing solution,tap water,and distilled water to prevent contamination and corrosion of the colorimetric bottle

  • Before inserting the colorimetric bottle into the colorimetric tank,the surface of the colorimetric tank,the surface of the colorimetric bottle must be wiped with soft paper or soft cloth,and the transparent surface of the colorimetric bottle must not be taken directly by hand to avoid affecting the measurement results

  • After the colorimetric tank,it needs to be placed for 10 to 30 seconds before measurement

  • when using devices such as pipettes,filters,digital titrators,etc.to absorb different solutions,the pipette suction head must be replaced, and the filter and digital titrator must be cleaned before absorbing another solution

  • If the reading value is unstable, the possible reasons are: (1) there is water on the outside wall of the colorimeter bottle, wipe the outside wall of the colorimeter bottle with filter paper, (2) the solution in the bottle is not stationary after the colorimeter bottle is placed in the colorimeter tank, (3) the battery voltage is insufficient (display L), the battery should be replaced


Testing instrument

Chlorine detection instrument can detect the concentration of hypochlorous acid produced by the equipment,which is accurate and has small error

Testing instrument


Leave your messages

Related products

Popular products

We are the source manufacturer of electrolytic water industry for 26 years, electrolytic water equipment has been updated to V4.0, and there are 16 agents around the world.
x